Clinical application of robotic three-dimensional targeting system in CT-guided lung biopsy |
CHEN Xiaoxia1, LIU Hao1, ZHAO Guoquan1, WANG Ti1, HOU Wenwen1, WANG Guisheng1, WANG Chuhan2 |
1.Radiology Department of General Hospital of Chinese People’s Armed Police Force,Beijing 100039,China; 2.Capital Medical Univarsity School of Bosic Medical Sciences,Beijing 100069,China |
|
|
Abstract Objective To explore the clinical value of MAXIOTM robotic three-dimensional targeting system in lung puncture biopsy guided by CT.Methods Between August and September 2017, 40 patients were randomly divided into the experimental group and control group. The experimental group adopted the MAXIOTM robotic three-dimensional targeting positioning system in the CT guided descending pulmonary puncture biopsy, while the control group was guided by the traditional grid locator to conduct the puncture biopsy. The average length of operation, one-time puncture success rate, scanning frequency, complication rate and biopsy pathological diagnosis rate were compared between the two groups.Results The average length of operation of lung puncture in the experimental group and control group was (24.0±3.9) min and (27.3±4.1) min respectively, so the difference between the two groups was not statistically significant(t=-1.91,P>0.05). The one-time puncture success rate of the experimental group and control group was 75% (15/20) and 25% (5/20)respectively, suggesting that the difference between the two groups was statistically significant(χ2=8.10,P<0.05). The average scanning times of the experimental group and the control group were (4.1±0.5)times and (6.0±1.3)times respectively, and the difference between the two groups was statistically significant(t=-4.65,P<0.05). The incidence of complications was 10% (2/20) and 40% (8/20) respectively in the experimental group and the control group so that the difference between the two groups was statistically significant(χ2=4.68, P<0.05). The positive rate of test in the experimental group and the control group was 100% (20/20) and 90% (18/20) respectively, so the difference between the two groups was not statistically significant (χ2=2.05,P>0.05).Conclusions The MAXIOTM robotic 3D targeting system can accurately plan the puncture path before CT-guided lung biopsy. The real-time stereotactic operation of the robot arm can improve the success rate of disposable puncture, thus reducing the times of scans, radiation dose and complications.
|
Received: 05 April 2018
|
|
|
|
|
[1] |
Haaga J R, Alfidi R J.Precise biopsy localization by computed tomography[J]. Radiology, 1976, 118(3): 603-607.
|
[2] |
金玉晶. 低剂量螺旋CT肺癌筛查检出肺结节的处理及进展[J].癌症进展,2017,1(4):349-353.
|
[3] |
李家开,于淼,肖越勇,等. CT透视引导下经皮放射性125I例子永久植入治疗肺恶性肿瘤[J].中国医学影像技术,2012, 28(9):1664-1669.
|
[4] |
胡佳寿,张立安,孙金凤,等.自动活检枪与CT-pinpoint三维激光定位导引系统的临床应用[J].中国医学影像学杂志,2006, 14(3):184-188.
|
[5] |
杨杰,肖越勇,张肖,等. 电磁导航系统在CT引导下经皮穿刺肺活检术中的应用[J].中国介入影像与治疗学,2012, 9(3):172-177.
|
[6] |
Krucker J, Xu S, Venkatesan A, et al. Clinical utility of resl-time fusion guidance for biopsy and ablation[J]. J Vasc Interv Radiol, 2011, 22(4):515-520.
|
[7] |
任伟,李茹恬,闫婧,等.改良CT和激光双引导经皮肺穿刺活检在肺癌精确诊断中的应用[J].中华肿瘤防治杂志,2017, 24(19):1373-1377.
|
[8] |
金雪云,嵇爱平,刘国平,等,CT引导下经皮肺穿刺活检在肺外周病变中的诊断价值[J]. 当代医学,2014, 10(28):72-73.
|
[9] |
Nakatani M, Tanigawa N, Kariya S, et al. Analysis of factors influencing accuracy and complications in CT-guided lung biopsy[J]. Minim Invasive Ther Allied Technol, 2012, 21(6): 415-422.
|
[10] |
Wu C C, Macher M M, Shepard J A.Complications of CT guided percutaneous needle biopsy of the chest: prevention and management[J]. AJR Am J Roentgenol, 2011, 196(6): W678-682.
|
[11] |
Wang Y, Jiang F, Tan X, et al. CT-guided percutaneous transthoracic needle biopsy for paramediastinal and nonparamediasinal lung lesion: diagnostic yield and complications in 1848 patients[J]. Medicine, 2016, 95(31):e4460.
|
[12] |
Yoshimatsu R, Yamagami T, Kato T, et al. Percutaneous needle biopsy of lung nodules under CT fluoroscopic guidance with use of the “I-I device”[J]. Br Radiol, 2008, 81(962):107-112.
|
[13] |
Lee S M, Park C M,Lee K H, et al, C-arm cone-beam CT-guided percutaneous transthoracic needle biopsy of lung nodules: clinical experience in 1108 patients[J]. Radiology, 2013, 271(1): 291-300.
|
[14] |
杨清杰,胡蒙,郭明. 新型三维穿刺定位引导器在经皮肺穿刺活检术中的应用[J].中华微创外科杂志,2016, 16(11):1019-1022.
|
|
|
|